The goal of this study was to develop a new implantable transducer for measuring anterior cruciate ligament (ACL) graft tension postoperatively in patients who have undergone ACL reconstructive surgery. A unique approach was taken of integrating the transducer into a femoral fixation device. To devise a practical in vivo calibration protocol for the fixation device transducer (FDT), several hypotheses were investigated: (1) The use of a cable versus the actual graft as the means for applying load to the FDT during calibration has no significant effect on the accuracy of the FDT tension measurements; (2) the number of flexion angles at which the device is calibrated has no significant effect on the accuracy of the FDT measurements; (3) the friction between the graft and femoral tunnel has no significant effect on measurement accuracy. To provide data for testing these hypotheses, the FDT was first calibrated with both a cable and a graft over the full range of flexion. Then graft tension was measured simultaneously with both the FDT on the femoral side and load cells, which were connected to the graft on the tibial side, as five cadaver knees were loaded externally. Measurements were made with both standard and overdrilled tunnels. The error in the FDT tension measurements was the difference between the graft tension measured by the FDT and the load cells. Results of the statistical analyses showed that neither the means of applying the calibration load, the number of flexion angles used for calibration, nor the tunnel size had a significant effect on the accuracy of the FDT. Thus a cable may be used instead of the graft to transmit loads to the FDT during calibration, thus simplifying the procedure. Accurate calibration requires data from just three flexion angles of 0, 45, and 90 deg and a curve fit to obtain a calibration curve over a continuous range of flexion within the limits of this angle group. Since friction did not adversely affect the measurement accuracy of the FDT, the femoral tunnel can be drilled to match the diameter of the graft and does not need to be overdrilled. Following these procedures, the error in measuring graft tension with the FDT averages less than 10 percent relative to a full-scale load of 257 N.

1.
Anderson
A. F.
, and
Lipscomb
A. B.
, “
Analysis of rehabilitation techniques after anterior cruciate reconstruction
,”
Am. J. Sports Med.
,
17
:
154
160
,
1989
.
2.
Bach
J. M.
, and
Hull
M. L.
, “
A new load application system for in vitro study of ligamentous injuries to the human knee joint
,”
ASME JOURNAL OF BIOMECHANICAL ENGINEERING
,
117
:
373
382
,
1995
.
3.
Ballock
R. T.
,
Woo
S. L.-Y.
,
Lyon
R. M.
,
Hollis
J. M.
,
Akeson
W. H.
, “
Use of patellar tendon autograft for anterior cruciate ligament reconstruction in the rabbit: A long-term histologic and biomechanical study
,”
J. Orthop. Res.
,
7
:
474
485
,
1989
.
4.
Barry
D.
, and
Ahmed
A. M.
, “
Design and performance of a modified buckle transducer for the measurement of ligament tension
,”
ASME JOURNAL OF BIOMECHANICAL ENGINEERING
,
108
:
149
152
,
1986
.
5.
Berns
G. S.
,
Hull
M. L.
,
Patterson
H. A.
, “
Implementation of a five degree of freedom automated system to determine knee flexibility in vitro
,”
ASME JOURNAL OF BIOMECHANICAL ENGINEERING
,
112
:
392
400
,
1990
.
6.
Beynnon
B.
,
Yu
J.
,
Huston
D.
,
Fleming
B.
,
Johnson
R.
,
Haugh
L.
,
Pope
M. H.
, “
A sagittal plane model of the knee and cruciate ligaments with application of a sensitivity analysis
,”
ASME JOURNAL OF BIOMECHANICAL ENGINEERING
,
118
:
227
238
,
1996
.
7.
Butler
D. L.
,
Grood
E. S.
,
Noyes
F. R.
,
Olmstead
M. L.
,
Hohn
R. B.
,
Arnoczky
S. P.
,
Siegel
M. G.
, “
Mechanical properties of primate vascularized vs. nonvascularized patellar tendon grafts; Changes over time
,”
J. Orthop. Res.
,
7
:
68
79
,
1989
.
8.
Flandry
F.
,
Terry
G. C.
,
Montgomery
R. D.
,
Kester
M. A.
,
Madsen
N.
, “
Accuracy of clinical isometry and preload testing during anterior cruciate ligament reconstruction
,”
Clin. Orthop.
,
279
:
214
222
,
1992
.
9.
Fleming, B. C., Good, L., Peura, G. D., Beynnon, B. D., “Measurement of anterior cruciate ligament (ACL) graft loads using an arthroscopic implantable force probe,” Trans. of the 42nd Annual Meeting of ORS, 21: 75, 1996.
10.
Glos
D. L.
,
Butler
D. L.
,
Grood
E. S.
,
Levy
M. S.
, “
In vitro evaluation of an implantable force transducer (IFP) in a patellar tendon model
,”
ASME JOURNAL OF BIOMECHANICAL ENGINEERING
,
115
:
335
342
,
1993
.
11.
Goss
B. C.
,
Hull
M. L.
,
Howell
S. M.
, “
Injury mechanisms in anterior cruciate ligament grafts from intercondylar roof impingement during passive knee extension
,”
J. Orthop. Res.
,
15
:
263
268
,
1997
.
12.
Graf, B., “Isometric placement of substitutes for the anterior cruciate ligament,” in: The Anterior Cruciate Deficient Knee: New Concepts in Ligament Repair, D. W. Jackson and D. Drez, Jr., eds., St. Louis, CV Mosby, 1987, pp. 102–113.
13.
Grood
E. S.
, and
Suntay
W. J.
, “
A joint coordinate system for the clinical description of three dimensional motions: application to the knee
,”
ASME JOURNAL OF BIOMECHANICAL ENGINEERING
,
105
:
136
144
,
1983
.
14.
Herzog
W.
,
Archambault
J. M.
,
Leonard
T. R.
,
Nguyen
H. K.
, “
Evaluation of the implantable force transducer for chronic tendon-force recordings
,”
J. Biomech.
,
29
:
103
109
,
1996
.
15.
Holden
J. P.
,
Grood
E. S.
,
Korvick
D. L.
,
Cummings
J. F.
,
Butler
D. L.
,
Bylski-Austrow
D. I.
, “
In vivo forces in the anterior cruciate ligament: Direct measurements during walking and trotting in a quadruped
,”
J. Biomech.
,
27
:
517
526
,
1994
.
16.
Holden
J. P.
,
Grood
E. S.
,
Cummings
J. F.
, “
Factors affecting sensitivity of a transducer for measuring anterior cruciate ligament force
,”
J. Biomech.
,
28
:
99
102
,
1995
.
17.
Howell
S. M.
, “
Arthroscopically assisted technique for preventing roof impingement of an anterior cruciate ligament graft illustrated by the use of an autogenous double-looped semitendinosus and gracilis graft
,”
Oper. Tech. Sports Med.
,
1
:
58
65
,
1993
.
18.
Howell
S. M.
, and
Gottlieb
J.
, “
Endoscopic fixation of a double-looped semitendinosus and gracilis ACL graft using bone mulch screw
,”
Oper. Tech. Orthop.
,
6
:
152
160
,
1996
.
19.
Lewis
J. L.
,
Lew
W. D.
,
Schmidt
J.
, “
A note on the application and evaluation of the buckle transducer for knee ligament force measurement
,”
ASME JOURNAL OF BIOMECHANICAL ENGINEERING
,
104
:
125
128
,
1982
.
20.
McKee, E. L., Lindsey, D., Hull, M. L., Howell, S. M., “Telemetry system for monitoring anterior cruciate ligament graft forces in vivo,” to appear in Med. Biol. Eng. Comp., 1998.
21.
Ray, J. D., Feder, S. M., Butler, D. L., Korvick, D. L., “Simultaneous in vivo measurement of tendon force and strain in the goat model,” Proc. 39th Annual Meeting of the ORS, 18 153, 1993.
22.
Rodeo
S. A.
,
Arnoczky
S. P.
,
Torzilli
P. A.
,
Hidaka
C.
,
Warren
R. F.
, “
Tendon healing in a bone tunnel: a biomechanical and histological study in the dog
,”
J. Bone Joint. Surg.
,
75A
:
1795
1803
,
1993
.
23.
Tohyama, H., Beynnon, B. D., Fleming, B. C., Peura, G. C., Johnson, R. J., Pope, M. H., “Evaluation of the arthroscopic force probe; a study of the factors that effect the measurement of tendon and ligament force in-vivo,” Trans, of 40th Annual Meeting of ORS, 19: 644, 1994.
24.
Wallace
M.
,
Howell
S. M.
,
Hull
M. L.
, “
In vivo tensile behavior of a four-bundled hamstrings graft as a replacement for the anterior cruciate ligament
,”
J. Orthop. Res.
,
15
:
539
545
,
1997
.
25.
Xu
W. S.
,
Butler
D. L.
,
Stouffer
D. C.
,
Grood
E. S.
,
Glos
D. L.
, “
Theoretical analysis of an implantable force transducer for tendon and ligament structures
,”
ASME JOURNAL OF BIOMECHANICAL ENGINEERING
,
114
:
170
177
,
1992
.
This content is only available via PDF.
You do not currently have access to this content.