The computational models developed to evaluate the hip joint performance usually neglect the presence of the pelvic bone. However, deformation depends on the stiffness of the underlying bone, and thus, the inclusion of the pelvic bone in the model influences the computed contact pressure and wear. This work discusses the influence of the pelvic bone, and how it depends on the acetabular component stiffness. It was modeled as two different polyethylene acetabular cups, considering or not a metal-backing for both 28 mm and 32 mm diametric cups. Two finite element models are developed, considering either the acetabular component rigidly fixed or attached to the deformable bone. Results present 28% and 42% difference on the contact pressure for a polyethylene cup without metal-backing when the support conditions are changed, for the 28 mm and 32 mm cups, respectively. Linear wear results present 21% and 31% difference for the same type of cups of 28 mm and 32 mm, correspondingly. The numerical results obtained in the present work show that to model the pelvic bone of the patient with a metal-backed cup did not greatly affect contact pressures and linear wear. However, when a total hip replacement is performed with an all-polyethylene acetabular cup, the presence of the pelvic bone in the model has a major influence.

1.
Maxian
,
T. A.
,
Brown
,
T. D.
,
Pedersen
,
D. R.
, and
Callaghan
,
J. J.
, 1996, “
A Sliding-Distance-Coupled Finite Element Formulation for Polyethylene Wear in Total Hip Arthroplasty
,”
J. Biomech.
0021-9290,
29
(
5
), pp.
687
692
.
2.
Maxian
,
T. A.
,
Brown
,
T. D.
,
Pedersen
,
D. R.
, and
Callaghan
,
J. J.
, 1996, “
Adaptative Finite Element Modelling of Long-Term Polyethylene Wear in Total Hip Arthroplasty
,”
J. Orthop. Res.
0736-0266,
14
, pp.
668
675
.
3.
Fialho
,
J. C.
,
Fernandes
,
P. R.
,
Eça
,
L.
, and
Folgado
,
J.
, 2007, “
Computational Hip Joint Simulator for Wear and Heat Generation
,”
J. Biomech.
0021-9290,
40
, pp.
2358
2366
.
4.
Ong
,
K. L.
,
Lehman
,
J.
,
Notz
,
W. I.
,
Santner
,
T. J.
, and
Bartel
,
D. L.
, 2006, “
Acetabular Cup Geometry and Bone-Implant Interference Have More Influence on Initial Periprosthetic Joint Space Than Joint Loading and Surgical Cup Insertion
,”
J. Biomech. Eng.
0148-0731,
128
, pp.
169
175
.
5.
Dalstra
,
M.
,
Huiskes
,
R.
,
Odgaard
,
A.
, and
van Erning
,
L.
, 1993, “
Mechanical and Textural Properties of Pelvic Trabecular Bone
,”
J. Biomech.
0021-9290,
26
, pp.
523
535
.
6.
Dalstra
,
M.
,
Huiskes
,
R.
, and
van Erning
,
L.
, 1995, “
Development and Validation of a Three Dimensional Finite Element Model of the Pelvic Bone
,”
J. Biomech. Eng.
0148-0731,
117
, pp.
272
278
.
7.
Andreykiv
,
A.
,
Prendergast
,
P. J.
,
van Keulen
,
F.
,
Swieszkowski
,
W.
, and
Rozing
,
P. M.
, 2005, “
Bone Ingrowth Simulation for a Concept Glenoid Component Design
,”
J. Biomech.
0021-9290,
38
, pp.
1023
1033
.
8.
Edidin
,
A. A.
, and
Kurtz
,
S. M.
, 2000, “
The Required Mechanical Properties for Hip and Knees Components
,” Ph.D. thesis, Drexel University and Exponent, Inc., Philadelphia, PA.
9.
Hopkins
,
A. R.
,
Hansen
,
U. N.
,
Amis
,
A. A.
,
Knight
,
L.
,
Taylor
,
M.
,
Levy
,
O.
, and
Copeland
,
S. A.
, 2007, “
Wear in the Prosthetic Shoulder: Association With Design Parameters
,”
J. Biomech. Eng.
0148-0731,
129
(
2
), pp.
223
230
.
10.
Bergmann
,
G.
,
Deuretzbacher
,
G.
,
Heller
,
M.
,
Graichen
,
F.
,
Rohlmann
,
A.
,
Strauss
,
J.
, and
Duda
,
G. N.
, 2001, “
Hip Contact Forces and Gait Patterns from Routine Activities
,”
J. Biomech.
0021-9290,
34
(
7
), pp.
859
871
.
You do not currently have access to this content.