The computational models developed to evaluate the hip joint performance usually neglect the presence of the pelvic bone. However, deformation depends on the stiffness of the underlying bone, and thus, the inclusion of the pelvic bone in the model influences the computed contact pressure and wear. This work discusses the influence of the pelvic bone, and how it depends on the acetabular component stiffness. It was modeled as two different polyethylene acetabular cups, considering or not a metal-backing for both 28 mm and 32 mm diametric cups. Two finite element models are developed, considering either the acetabular component rigidly fixed or attached to the deformable bone. Results present 28% and 42% difference on the contact pressure for a polyethylene cup without metal-backing when the support conditions are changed, for the 28 mm and 32 mm cups, respectively. Linear wear results present 21% and 31% difference for the same type of cups of 28 mm and 32 mm, correspondingly. The numerical results obtained in the present work show that to model the pelvic bone of the patient with a metal-backed cup did not greatly affect contact pressures and linear wear. However, when a total hip replacement is performed with an all-polyethylene acetabular cup, the presence of the pelvic bone in the model has a major influence.
Skip Nav Destination
e-mail: saraloureirobarreto@gmail.com
e-mail: jfolgado@dem.ist.utl.pt
e-mail: prfernan@dem.ist.utl.pt
e-mail: jac.monteiro@hsm.min-saude.pt
Article navigation
May 2010
Technical Briefs
The Influence of the Pelvic Bone on the Computational Results of the Acetabular Component of a Total Hip Prosthesis
Sara Barreto,
Sara Barreto
IDMEC-Instituto Superior Técnico,
e-mail: saraloureirobarreto@gmail.com
Technical University of Lisbon
, Avenida Rovisco Pais, 1049-001 Lisbon, Portugal
Search for other works by this author on:
João Folgado,
João Folgado
IDMEC-Instituto Superior Técnico,
e-mail: jfolgado@dem.ist.utl.pt
Technical University of Lisbon
, Avenida Rovisco Pais, 1049-001 Lisbon, Portugal
Search for other works by this author on:
Paulo R. Fernandes,
Paulo R. Fernandes
IDMEC-Instituto Superior Técnico,
e-mail: prfernan@dem.ist.utl.pt
Technical University of Lisbon
, Avenida Rovisco Pais, 1049-001 Lisbon, Portugal
Search for other works by this author on:
Jacinto Monteiro
Jacinto Monteiro
Department of Orthopaedics, Hospital Santa Maria—Faculty of Medicine,
e-mail: jac.monteiro@hsm.min-saude.pt
University of Lisbon
, Avenida Professor Egas Moniz, 1649-028 Lisbon, Portugal
Search for other works by this author on:
Sara Barreto
IDMEC-Instituto Superior Técnico,
Technical University of Lisbon
, Avenida Rovisco Pais, 1049-001 Lisbon, Portugale-mail: saraloureirobarreto@gmail.com
João Folgado
IDMEC-Instituto Superior Técnico,
Technical University of Lisbon
, Avenida Rovisco Pais, 1049-001 Lisbon, Portugale-mail: jfolgado@dem.ist.utl.pt
Paulo R. Fernandes
IDMEC-Instituto Superior Técnico,
Technical University of Lisbon
, Avenida Rovisco Pais, 1049-001 Lisbon, Portugale-mail: prfernan@dem.ist.utl.pt
Jacinto Monteiro
Department of Orthopaedics, Hospital Santa Maria—Faculty of Medicine,
University of Lisbon
, Avenida Professor Egas Moniz, 1649-028 Lisbon, Portugale-mail: jac.monteiro@hsm.min-saude.pt
J Biomech Eng. May 2010, 132(5): 054503 (4 pages)
Published Online: March 29, 2010
Article history
Received:
November 12, 2009
Revised:
January 12, 2010
Posted:
January 18, 2010
Published:
March 29, 2010
Online:
March 29, 2010
Citation
Barreto, S., Folgado, J., Fernandes, P. R., and Monteiro, J. (March 29, 2010). "The Influence of the Pelvic Bone on the Computational Results of the Acetabular Component of a Total Hip Prosthesis." ASME. J Biomech Eng. May 2010; 132(5): 054503. https://doi.org/10.1115/1.4001031
Download citation file:
Get Email Alerts
Simultaneous Prediction of Multiple Unmeasured Muscle Activations Through Synergy Extrapolation
J Biomech Eng (March 2025)
Related Articles
Simulation of Physiological Loading in Total Hip Replacements
J Biomech Eng (August,2006)
Design Concept for a Total Knee Replacement with Condylar Guiding Features
J. Med. Devices (June,2011)
A New External Fixator Design for Femoral Fracture Reduction
J. Med. Devices (June,2008)
Bending Vibrations of the Femur and the Oscillatory Behavior of a Cemented Femoral Hip Endoprosthesis
J Biomech Eng (August,2000)
Related Proceedings Papers
Related Chapters
Novel and Efficient Mathematical and Computational Methods for the Analysis and Architecting of Ultralight Cellular Materials and their Macrostructural Responses
Advances in Computers and Information in Engineering Research, Volume 2
Application of Non-Linear Elastic Wave Spectroscopy (NEWS) to In Vitro Damage Assessment in Cortical Bone
Biomedical Applications of Vibration and Acoustics in Imaging and Characterizations
Computer Aided Oracle Bone Inscriptions Textual Research Based on Ontology
International Conference on Mechanical and Electrical Technology, 3rd, (ICMET-China 2011), Volumes 1–3