Abstract
During STR analysis on the 310 Genetic Analyzer, retarded migration of GS500ROX size standards and alleles in some samples was observed. The contribution of reagents, capillary and performance optimized polymer POP 4 to the observed anomaly was experimentally eliminated. Variation in electrophoresis temperature between 55°C and 65°C did not alter the rate of migration of GX500ROX size standard and sample alleles. An eroded connector for the cathode mounted on the heat plate assembly caused the abnormal migration. Hence, it is important to verify the mobility of all fragments in the size standard for each sample to avoid any erroneous allele calls by the automated data analysis software.
Issue Section:
Research Papers
References
1.
Edward
A
, Hammond
H A
, Lin
J
, Caskey
C T
, Chakraborty
R
. Genetic variation at five trimetric and tetrametric tandem repeat loci in four human population groups
. Genomics
0888-7543 1992
;12
:241
-53
.2.
Fregeau
C J
, Fourney
R M
. DNA typing with fluoroscently tagged short tandem repeats: a sensitive and accurate approach to human identification
. Biotechniques
1993
;15
:100
-19
.3.
Kimpton
C P
, Gill
P
, Walton
A
, Milligan
E S
, Adams
M
. Automated DNA profiling employing multiplex amplification of short tandem repeat loci
. PCR Meth and Applic
1993
;3
:13
-22
.4.
Urquhart
A
, Oldroyd
N J
, Kimpton
C P
, Gill
P
. Highly discriminating heptaples short tandem repeat PCR system for forensic identification
. Biotechniques
1995
;18
:116
-21
.5.
Lins
A M
, Micka
K A
, Sprecher
C J
, Taylor
J A
, Bachner
J W
, Rabbach
D R
, et al. Development and population study of an eight-locus short tandem repeat (STR) multiplex system
. J Forensic Sci
1998
;43
:1168
-80
.
This content is only available via PDF.
All rights reserved. This material may not be reproduced or copied, in whole or in part, in any printed, mechanical, electronic, film, or other distribution and storage media, without the written consent of ASTM International.
You do not currently have access to this content.