A weight bearing indicator (WBI) has been developed and is undergoing clinical trials and commercialization. The most recent version of the device includes load range indication ability. The WBI is installed in a patient’s medical appliance (e.g. cam walker boot, surgical shoe) and allows patients recovering from lower extremity injuries to self monitor their rehabilitation. It is a simple and inexpensive mechanical device that utilizes the reversible buckling phenomena of a snap dome to provide a tactile ‘click’ and an audible ‘snap’ when a specified ground load on the bottom of a patient’s foot, which has been prescribed by an orthopedist or podiatrist as the partial weight bearing (PWB) upper limit, has been met or exceeded. A clinical study involving 20 subjects showed that PWB compliance improved when comparing the WBI device to standard of care approaches such as verbal instruction and the weight scale method. Specifically, PWB compliance improved from 58% and 45% body weight (BW) for verbal instruction and weight scale method, respectively, to 33% BW for the WBI device. In addition, PWB compliance using the WBI device was not dependent on subject weight like the standard of care approaches. Sometimes patients are instructed not only to avoid exceeding an upper load limit, but also to maintain their limb load above a lower limit as well. A newer version of the WBI has been developed that provides two distinct clicks to a patient corresponding to an upper and lower load limits.
Skip Nav Destination
Article navigation
Design Of Medical Devices Conference Abstracts
Progress on Clinical Trials and Product Development of a Simple, Inexpensive Feedback Device for Patients Requiring Partial Weight Bearing
Daniel Walczyk,
Daniel Walczyk
Rensselaer Polytechnic Institute
, Troy, NY and SUNY Upstate Medical University
, Syracuse, NY USA
Search for other works by this author on:
Nathaniel Ordway,
Nathaniel Ordway
Rensselaer Polytechnic Institute
, Troy, NY and SUNY Upstate Medical University
, Syracuse, NY USA
Search for other works by this author on:
Kenneth Mann,
Kenneth Mann
Rensselaer Polytechnic Institute
, Troy, NY and SUNY Upstate Medical University
, Syracuse, NY USA
Search for other works by this author on:
William Ziomek
William Ziomek
Rensselaer Polytechnic Institute
, Troy, NY and SUNY Upstate Medical University
, Syracuse, NY USA
Search for other works by this author on:
Daniel Walczyk
Rensselaer Polytechnic Institute
, Troy, NY and SUNY Upstate Medical University
, Syracuse, NY USA
Nathaniel Ordway
Rensselaer Polytechnic Institute
, Troy, NY and SUNY Upstate Medical University
, Syracuse, NY USA
Kenneth Mann
Rensselaer Polytechnic Institute
, Troy, NY and SUNY Upstate Medical University
, Syracuse, NY USA
William Ziomek
Rensselaer Polytechnic Institute
, Troy, NY and SUNY Upstate Medical University
, Syracuse, NY USAJ. Med. Devices. Jun 2008, 2(2): 027528 (1 pages)
Published Online: June 12, 2008
Article history
Published:
June 12, 2008
Citation
Walczyk, D., Ordway, N., Mann, K., and Ziomek, W. (June 12, 2008). "Progress on Clinical Trials and Product Development of a Simple, Inexpensive Feedback Device for Patients Requiring Partial Weight Bearing." ASME. J. Med. Devices. June 2008; 2(2): 027528. https://doi.org/10.1115/1.2936203
Download citation file:
591
Views
Get Email Alerts
Cited By
A Better Way to Visualize the Cervix: A Low-Cost Cervical Cancer Screening Device for LMIC Settings
J. Med. Devices (April 2025)
Comparative Survey-Based Study of Noninvasive Saliva Collection Devices
J. Med. Devices (April 2025)
Related Articles
Medtech Design in Interdisciplinary Clinical Innovation Teams
J. Med. Devices (June,2009)
The Design of a Cell-Phone Based Balance-Training Device
J. Med. Devices (June,2009)
An Inexpensive Weight Bearing Indicator Used for Rehabilitation of Patients With Lower Extremity Injuries
J. Med. Devices (March,2007)
Vibrotactile Balance Rehabilitation Gait Assist Device
J. Med. Devices (June,2009)
Related Proceedings Papers
Related Chapters
Vibration Analysis of the Seated Human Body in Vertical Direction
International Conference on Computer Technology and Development, 3rd (ICCTD 2011)
IMPLICATIONS OF LIMIT STATE DESIGN CHECKS FOR GEOTECHNICAL LOADS
Pipeline Integrity Management Under Geohazard Conditions (PIMG)
Mechanics of Long Beam Columns
Mechanics of Drillstrings and Marine Risers